Eligibility of rotor assembly for duty-free tariff treatment under the Nairobi Protocol; 9817.00.96; specially designed or adapted for the handicapped; parts
No revocation or modification of this ruling appears in the Customs Bulletin issues read here (Nov 6, 2024 through Nov 18, 2026), where CBP publishes most such actions. That is not confirmation that the ruling is still in effect. Confirm its status with CBP before relying on it (ruling on CBP CROSS).
Ruling text
This is in response to your letter dated September 24, 2018, on behalf of Moog, Inc. (Moog), which requests a binding ruling regarding the eligibility of an imported rotor assembly for duty-free treatment under subheading 9817.00.96, Harmonized Tariff Schedule of the United States (HTSUS). FACTS: The article under consideration is a rotor assembly.
You state that this rotor assembly is specially designed for an electric motor that powers the compressor of a portable oxygen concentrator (POC), a medical device that supplies supplementary oxygen to its users. Moog specially produces this rotor assembly for Inogen, a manufacturer and seller of POCs. Inogen custom orders this rotor assembly from Moog with specific requirements for the design and manufacture of the rotor assembly for use with its POCs.
The rotor assembly is manufactured in China and imported into the United States for processing prior to being used in the POC. As imported, the instant rotor assembly has certain physical features that differentiate it from Moog’s standard catalog rotors. While Moog’s standard catalog rotors are designed with a front shaft extension, the instant rotor features an additional rear shaft extension.
The dual-shaft design of the instant rotor allows it to maintain symmetry, which is required to operate the dual-piston function of the compressor in the POC. The front shaft extension operates one piston in the compressor, while the rear shaft extension operates the other. The tight tolerances and dimensions of the shaft extensions are specific to this application. In addition, you state that the shaft material used in the rotor assembly is unique to its design.
For the instant rotor, precipitation-hardening stainless steel is used for the shaft material, whereas standard rotors use martensitic stainless steel. Precipitation hardening is a process by which the impurities in the crystalline metal are rectified by application of a chemical solution and specific heating regime. The benefit of this process is increased durability or wear resistance, the ability to undergo high tensile and yield stress, and improved machinability.
The enhanced machinability of the precipitation-hardening steel allows the shaft to meet the tight tolerances required for its use in a POC. The improved characteristics of the metal provide the added quality and reliability required for use in a U.S. Food and Drug Administration (FDA) regulated medical device. As the rotor assembly at issue must be used in an oxygen-supplying medical device, Inogen requires higher quality components and processes.
Unlike standard rotor assemblies, the instant rotor assembly uses primer at the bond joint of the rotor core to the shaft. When being used in a POC, a slip of the rotor core on the shaft would result in a non-functioning motor and compressor, causing a loss of concentrated oxygen to the user. This risk is mitigated by specially machined glue grooves and primer where the shaft and rotor core meet. This results in a stronger bond between the components and a lower likelihood of failure.
The text above is an excerpt. The official ruling, with its letterhead, signature and any attachments, is on CBP CROSS (H301481).