The tariff classification of the CHIPSCANNER 150 Two-HS from Germany
No revocation or modification of this ruling appears in the Customs Bulletin issues read here (Nov 6, 2024 through Nov 18, 2026), where CBP publishes most such actions. That is not confirmation that the ruling is still in effect. Confirm its status with CBP before relying on it (ruling on CBP CROSS).
Key passages, quoted from the ruling
- Accordingly, the applicable subheading for the CHIPSCANNER 150 Two-HS will be 9031.80.8085, HTSUS, which provides for “Measuring or checking instruments, appliances and machines, not specified or included elsewhere in this chapter; profile projectors; parts and accessories thereof: Other instruments, appliances and machines: Other: Other.” The general rate of duty will be free.
Ruling text
In your letter dated April 16, 2025, you requested a tariff classification ruling. Descriptive literature was provided for our review. The item under consideration is the CHIPSCANNER 150 Two-HS, which is an advanced, ultra-high-precision, large-area Scanning Electron Microscope (SEM) imaging system specifically designed for semiconductor metrology, inspection, and design verification/reverse engineering applications.
It is a complete, fully integrated system comprising a specialized electron optical column, laser interferometer stage, high-vacuum system, control electronics, and sophisticated software for image acquisition and analysis. The electron optical column has a Schottky thermionic field emission electron source (ZrO/W filament), an electromagnetic/electrostatic objective lens system with beam booster, a stigmator and aperture system, and a digital beam deflection system.
The secondary electron detectors include an in-lens secondary electron detector for imaging up to 20 kV, an off-axis Everhart-Thornley secondary electron detector (SE2) with 20 ns pixel response, and an optional backscattered electron detector. The laser interferometer stage system includes a high precision X-Y-Z stage with 150mm by 150mm travel range, a laser interferometer position sensing in X and Y planes with 1 nm resolution, and DC motors with piezo actuators for fine positioning.
The sample handling and chamber include an automated load-lock system for sample exchange, a universal sample holder system capable of handling wafers up to 8-inch diameter, an internal infrared camera system for sample navigation, and a three-point sample leveling mechanism with piezo actuators. The high vacuum system includes an oil-free high vacuum system with turbo-molecular pump, a dry vacuum pre-pump, and an ion getter pump for ultra-high vacuum (UHV) maintenance in gun region.
Finally, the control electronics and software include a 50 MHz field-programmable gate array (FPGA)-controlled high-speed scan generator, Windows 10 64-bit control system with multi-user interface, and advanced imaging software with computer aided design (CAD) navigation capabilities. The device measures 2100mm by 1400mm by 2100mm and weighs approximately 1700 kg.
In your submission, you suggest the classification of the CHIPSCANNER 150 Two-HS in subheading 8486.20.0000, Harmonized Tariff Schedule of the United States (HTSUS), which provides for “Machines and apparatus of a kind used solely or principally for the manufacture of semiconductor boules or wafers, semiconductor devices, electronic integrated circuits or flat panel displays; machines and apparatus specified in note 11(C) to this chapter; parts and accessories: Machines and apparatus for the manufacture of semiconductor devices or of electronic integrated circuits.” We disagree.
The purpose and function of the CHIPSCANNER Two-HS is to inspect and perform metrology on semiconductor wafers and devices. Apparatus for measuring, checking, inspecting, and performing measurements are classified in Chapter 90, HTSUS.
The text above is an excerpt. The official ruling, with its letterhead, signature and any attachments, is on CBP CROSS (N347907).